Finden Sie Interpretive Ecg Machine, Ecg Maschine, Dreikanal-EKG auf Industry Directory, zuverlässige Hersteller / Lieferant / Fabrik aus China
neue Produkte
Lieferzeit: | 5 Tage |
---|
Modell: MSLPE03K
Items |
specification |
Lead |
standard 12 leads |
lead acquisition |
synchronously 12 leads |
Data acquisition |
18bit/1000Hz |
Record mode |
Manual mode; Auto 1; Auto 2; Auto 3 |
Filter |
AC, EMG,Drift Filter |
CMRR |
100dB(with AC filter) |
Input circuit |
Floating; Protection circuit against Defibrillator effect |
Input circuit current |
≤0.1μA |
Input Impedance |
≥2.5MΩ |
Patient current leakage |
<10μA |
Calibrating voltage |
1mV±2% |
Voltage Tolerance |
±500mV |
Time constant |
>3.2s |
Frequency response |
0.05 ~150Hz |
Noisy level |
≤15μVp-p |
Sensitivity valve |
≤20μV |
Inter-Channel Interference |
≤0.5mm |
Paper speed |
25,50mm/s(±3%) |
Sensitivity |
AUTO,2.5,5,10,20mm/mV |
Recording mode |
Thermal Array printing |
Paper specification |
50mmx20m,roll |
LCD display |
20x2 character LCD,display status. |
Safety classification |
IEC60601-1 class I, type CF |
Power supply |
AC : 110V/220V/85~265V,50/60Hz±1Hz;30VA |
DC :12V(1500mAh) rechargeable Ni-MH battery | |
Fuse |
AC extend T500mA/125V Φ5x20mm AC extend T250mA/250V Φ5x20mm AC extend T2.0A/250V Φ5x20mm |
Produktgruppe : EKG und EKG